Reference
ICH Q2(R2) and Q14 — working reference
The requirement matrix, what changed in the R2 revision, how the Q14 lifecycle fits around it, and the findings that come up most often in validation review.
ICH Q2(R2) — validation characteristics by procedure type
This matrix states which characteristics must be evaluated. It does not state numerical acceptance criteria — those are set and justified by the applicant.
| Characteristic | Identification | Impurities quantitative | Impurities limit test | Assay content / potency |
|---|---|---|---|---|
| Specificity / Selectivity | Yes | Yes | Yes | Yes |
| Accuracy | No | Yes | No | Yes |
| Precision (repeatability) | No | Yes | No | Yes |
| Intermediate precision | No | Yes | No | Yes |
| Linearity / Response function | No | Yes | No | Yes |
| Range | No | Yes | No | Yes |
| Detection limit (LOD) | No | May be needed | Yes | No |
| Quantitation limit (LOQ) | No | Yes | No | No |
| Robustness | Yes | Yes | Yes | Yes |
| Stability of solutions | May be needed | Yes | May be needed | Yes |
Dissolution, content uniformity, residual solvents, cleaning verification and bioanalytical procedures follow the same logic with purpose-specific expectations; select the purpose on the dashboard to see the applicable set. Bioanalytical procedures are governed by ICH M10 rather than Q2(R2).
Q2(R2) reorganised rather than rewrote. The most consequential changes for day-to-day work are these.
Range absorbed linearity. What used to be three separate characteristics — linearity, accuracy and precision — now sit under the working range, and the guideline speaks of a *response function* rather than assuming a straight line. A non-linear model is explicitly acceptable when justified. In practice this means the correlation coefficient carries less weight than it used to, and the residual plot and back-calculated accuracy carry more.
Stability of solutions became explicit. Previously an implicit expectation, now a named characteristic. The output is an operational claim — how long solutions may be held, under what condition — which belongs in the procedure text and constrains how long a routine sequence may be.
Multivariate procedures are addressed. Q2(R2) accommodates procedures such as NIR and Raman models, where validation of a model differs from validation of a single-analyte chromatographic measurement.
Prior knowledge is a legitimate input. In line with Q14, development data and platform knowledge can reduce the experimental burden at validation, provided they are documented and relevant.
What did not change: Q2(R2) still specifies WHICH characteristics must be evaluated, not what the numerical acceptance criteria should be. Criteria are yours to set and justify against the intended use.
Primary sources
- ICH Q2(R2) — Validation of Analytical Procedures (2023)
- ICH Q14 — Analytical Procedure Development (2023)
- ICH Q1A(R2) — Stability Testing of New Drug Substances and Products
- ICH Q1B — Photostability Testing
- ICH Q3A(R2) / Q3B(R2) — Impurities in New Drug Substances / Products
- ICH Q3C(R8) — Residual Solvents
- ICH M10 — Bioanalytical Method Validation and Study Sample Analysis
- USP <621> Chromatography; <1010> Analytical Data; <1225> Validation of Compendial Procedures; <1226> Verification of Compendial Procedures
- Ph. Eur. 2.2.46 — Chromatographic Separation Techniques
This reference summarises and interprets the guidelines for practical use. It is not a substitute for reading the current text of the guidelines themselves, and where this summary and a guideline differ, the guideline governs.